FDA · Life Science & Pharma: GxP - CSV & Data Integrity
Computer Software Assurance for Production and Quality Management System Software (FDA guidance)
Final CSV · Guidance Medical Device Item Related to CSV Last verified: not yet · Source last checked 3 Nov 2026
- Issued by
- FDA
- Category
- CSV
- Sub Category
- Guidance
- Legal weight
- Guidance & Q&A
- Reuse
- Public domain
FDA's guidance on checking software used in device production and quality systems. It asks for effort that matches the risk, not paperwork for its own sake.
How to read this page
- The line under the title explains in plain words what this item is for. The coloured label beside its status gives its Category, its main subject, and Sub Category, the type of document, in the same colours as the maps.
- Status comes from the latest dated event below, and each event links to its source.
- Lifecycle shows the steps so far, then any expected target. Targets are not facts.
- In plain English is our own summary, not the official text. Where we have read the document in detail, a fuller profile follows, with its context, scope, contents and key points.
- Related shows links to other documents. Select Compare to see two side by side.
Lifecycle
Solid steps have happened. A dashed step is expected, and it is not a fact. Only dated steps in our record appear here.
- Draft guidance published
- Final guidance published
- Updated and retitled for QMS software
In plain English
FDA's risk-based approach to assuring software used in production and quality systems, favouring critical thinking and unscripted testing where risk is lower. Final since September 2025 and updated in February 2026 to align with the QMSR.
Summary drafted with AI assistance. Not yet reviewed by the SME Reviewer. Read the official text before acting.
Context and history
Computer Software Assurance is FDA's risk-based approach to gaining confidence in software used in device production and quality systems. It aims to move effort from producing documents to critical thinking and testing where risk is highest. FDA issued a draft on 13 September 2022 and finalised it on 24 September 2025. This version, issued on 3 February 2026, supersedes that one, and updates the title and references to fit the Quality Management System Regulation, which took effect on 2 February 2026.
It builds on FDA's General Principles of Software Validation, and does not replace the basic idea of validating software for its intended use.
Who and what it applies to
Computers or automated data processing systems used as part of production or the quality management system for medical devices, for manufacturers subject to the QMSR. Drug manufacturers often read it for ideas, but it is written for devices.
Not covered: Software that is part of a device, and drug manufacturing, are not the subject of this guidance.
What it contains
- I to II Introduction and background
- Why FDA promotes this approach, and how it relates to the QMSR.
- III Scope
- What software is covered, and how this builds on FDA's Software Validation guidance.
- IV Definitions
- Terms used, including computer software assurance.
- V.A Risk framework
- Identify the software's intended use, determine a risk-based approach, determine appropriate assurance activities, and establish an appropriate record.
- V.B Electronic records
- Considerations for electronic records requirements.
- Appendix A Examples
- Worked examples, showing risk reasoning and assurance activities.
Key points
- Identify reasonably foreseeable software failures, and decide whether they pose a high process risk.
- Match assurance activities to that risk.
- Scripted testing is not the only option: unscripted testing can fit lower risk.
- The record should be enough to show the work, not more.
- Leverage supplier and other existing evidence where justified.
In practice
General reading, not advice. This is drafted with AI and may not fit your situation. It does not replace the official text, your own quality system or qualified advice, and an inspector may take a different view.
- Use the worked examples to reset how teams think about test documentation.
- Agree the approach with QA before a project starts, because practices vary.
- If you are outside devices, check how far your regulator accepts this thinking before relying on it.
Read from the source document on 1 Nov 2026. This is our own description, not the official text. Check the full dated history in the change log below.
Change log
Key dated events only, not a complete history. Status after is the status once that event happened. Event kinds explained
| Date | Event | Status after | Source |
|---|---|---|---|
| 3 Feb 2026 | Updated and retitled for QMS software | Final | Source for Updated and retitled for QMS software |
| 24 Sep 2025 | Final guidance published | Final | Source for Final guidance published |
| 13 Sep 2022 | Draft guidance published | Draft | Source for Draft guidance published |
Related
How other documents connect to this one. Links explained
- Complements FDA software validation· Compare with FDA software validation Unverified link
- Related to Part 11 scope guidance· Compare with Part 11 scope guidance Unverified link
- Related to ISPE GAMP 5, 2nd ed.· Compare with ISPE GAMP 5, 2nd ed. Unverified link
Official source
The official text always prevails over anything on this site.
Source: FDA. Public domain.
Stay updated
Newsletter signup opens at launch.